Print page

Metrology Service

Optical Homogeneity
Interferometry: 2D Maps with analysis

Measurement Wavelength: 632,8nm

Standard-Procedure:
  • Sample Diameter: ≤ 450mm,
    rectangular plates accordingly smaller
  • Sample Thickness: 22mm - 400mm
  • CA-Diameter: ≤ 430mm
    rectangular plates accordingly smaller
  • Accuracy: 2σ = 0,3 ppm

Polished Plate Technique
  • Sample Diameter: 135mm - 345 mm
  • Sample Thickness: 20 mm - 80 mm
  • CA Diameter:130 mm - 310 mm
  • Surface Quality: polished better than lambda / 2,
    no short term aberrations,
    tilt of about 4 arc minutes
  • Accuracy: 2σ = 0,05 ppm

Stiching-Technique
  • Sampe Diameter: ≤ 650 mm x 650mm
  • Sample Thickness: 20 mm - 240 mm
  • CA Diameter: ≤ 610 mm x 610mm
  • Accuracy: 2σ = 1 ppm
Stress induced Birefringence
2D map with analysis

  • Measurement Range: 0,01nm - 260 nm
  • Inspection Wavelength: 632,8 nm
  • Laser spot size: 1mm Diameter
  • Inspection Grid down to: 1mm x 1mm
  • Standard inspection grid: 10 mm x 10 mm
  • Accuracy: < 0,05 nm/cm
  • Sample Diameter: ≤ 450 mm x 450 mm
  • Sample Thickness: ≤ 385mm
  • CA: ≤ 450 mm x 450 mm
Transmission
190nm - 500nm: Transmissionsspectrum or single wavelength measurement

  • Measurement range: 190nm - 5000nm
  • Data: 1 Datapoint pernm
  • Accuracy: 0,5% Transmission
  • Sample Geometry:
    Thickness: min. 10mm
    Area: min.: 20mm x 20mm; max.: 100mm x 100mm
Bubble or Inclusion Inspection
Automated detection system down to 10µm defect size: 3D maps with analysis

  • Measurement range: ≥ 10µm
  • Accuracy: Defectsize ± 20%
  • Sample size: 500mm x 500mm x 140 mm
Density
  • Accuracy: 0,01 g/cm2
  • Sample size: sample volume ~ 1 cm3
Viscosity
  • Measurement range: 1000°C - 1500°C
  • Sample size: 55 mm x 5 mm, thickness 1 mm - 3 mm
  • Accuracy: lg 0,2 dPa s